Filtration Media Analyzer

<< Back to the Porometers Page

Description

The PMI Filtration Media Analyzer is designed to meet the through-pore measurement needs of the filtration industry. The Filtration Media Analyzer provides fully-automated analysis of bubble point, pore size distribution, mean pore size, gas, liquid, and Frazier permeability, filter integrity and hydro-head. Testing under compression, tension, and elevated temperature is also possible, thereby allowing the user to simulate actual operating conditions.

Principles of Operation

A fully wetted sample is placed in the sample chamber and the chamber is sealed. Gas is then allowed to flow into the chamber behind the sample. When the pressure reaches a point that can overcome the capillary action of the fluid within the pore (largest pore), the bubble point has been found. After determination of the bubble point, the pressure is increased until all pores are emptied of fluid, i.e., considered dry. Gas pressure and flow rates through dry samples are also measured.

Applications

PMI's Filtration Media Analyzer is designed to meet the needs of the filtration industry -- both in the R&D and quality control environments. The Filtration Media Analyzer characterizes a wide range of filter media, including microfiltration and ultrafiltration membranes.

Applicable industries are:

For more information about our applications please see our other technical papers.

Features

Specifications

Pore Size Range
0.013 - 500 microns
Permeability Range
1 x 10-10, 1 x 10-6 (Microflow in cc/sec/m/torr)
Sample Size
1.75" - 2.5" diameter
Pressure Range
0 - 500 psi
Pressurizing Gas
Clean, dry, compressed air or nonflammable, non-corrosive gas
Pressure Transducer Range
0 - 500 psi
Resolution
1 in 20,000
Accuracy
0.15% of reading
Mass Flow Transducer Range
10 cc/minute - 500 L/minute
Power Requirements
110/120 VAC, 50/60 Hz
Dimensions
30" H x 19" W x 18.5" D
Weight
100 lbs

For more information see our Porometers Brochure and Characterization of Pore Structure Brochure.

Back to the Top